June 25, 2024
Recently, the FDA updated a long-standing, decades old guidance on analytical test method validation based on revisions of the ICH Q2(R2) guidelines. Traditional test method validation requirements have been streamlined and altered to both provide flexibility on new types of multivariate and non-linear analytical test methods, and to focus on the validation-critical parameters that show that the method is reliable during routine use.
Validation Characteristics (ICH Q2 and ICH Q2(R1)) |
New Validation Characteristics (ICH Q2(R2)) |
---|---|
Specificity |
Specificity/Selectivity |
Linearity |
Range (includes handling of non-linear responses) |
Accuracy |
Accuracy/Precision |
Precision (repeatability, Intermediate precision, and reproducibility) |
|
Range |
|
Major Changes in Revision 2 |
|
Quantitation Limit |
1. Robustness and system suitability acceptance criteria determination now incorporated into method development |
Detection Limit |
2. Added criteria for Multivariate test methods |
Robustness |
3. Update method transfer requirements to require partial of full revalidation at receiving site |
What has not changed is the expectation that a test method is thoroughly developed and suitable for routine use. Test methods should have a defined purpose and measure some critical characteristic(s) of the drug substance or product. Validation protocols and pre-defined acceptance criteria should be written prior to validation. Analytical test method validations must be completed prior to NDA submission and prior to characterization and release of the pivotal clinical trial materials used in Phase III clinical studies.
Table 2: Newly Defined Analytical Test Method Ranges
Use of Analytical Procedure |
Low End of Reportable Range |
High End of Reportable Range |
---|---|---|
Assay of a Product |
80% of declared content or 80% of lower specification acceptance criterion |
120% of declared content or 120% of the upper specification acceptance criterion |
Potency |
Lowest specification acceptance criterion -20% |
Highest specification acceptance criterion +20% |
Content Uniformity |
70% of declared content |
130% of declared content |
Dissolution |
||
Immediate Release
|
|
|
Modified Release |
Lower limit of reportable range (as justified by the specification) or quantitation limit (QL), as appropriate |
|
Impurity1 |
Reporting threshold |
120% of the specification acceptance criterion |
Purity (as% area) |
80% of the lower specification acceptance criterion |
Upper specification acceptance criterion or 100% |
Tests should show an absence of interference from other substances and be specific to the target analyte. A lack of specificity can be compensated in some cases by utilizing a combination of two or more procedures. For example, denaturing gel electrophoresis may separate a protein monomer from a covalently linked dimer. However, non-covalently linked aggregates may need a secondary assay to quantitate. In some cases where the analytical technology is specific (NMR, MS, etc.), additional experimental studies may not be required if justified.
Required for: ID tests, Assay, Purity, and Impurity tests
The range of the assays must cover both the upper end and lower end of the specification limits and typically represent the highest and lowest sample concentrations where the procedure produced reliable results. For linear responses, the guidance is unchanged with predefined correlation coefficients, y-intercept, and slope of the regression lines still being required.
One significant change to the new guidance document is the incorporation of nonlinear responses. In these cases, a model or function describes the activity/concentration present and correlates it to the response. For example, many immunoassays show S-shaped responses rather than a linear response.
If the measure of the analyte is close to the lower range limit of the procedure (e.g., Impurities) the limit of quantitation and limit of detection should be established.
Required for: ID tests, Assay, Purity, and Impurity (both limit and quantitative) tests
Accuracy and precision of the assay can be evaluated independently or in a single study. Accuracy should be established across the range for the analytical procedure. Recovery studies of a known quantity of the analyte in the sample matrix are typically used. Triplicate samples at three different concentrations are analyzed at covering the range of the test method.
One change to the accuracy validation criteria is for multivariate analytical procedures. The test method should be evaluated for the root mean square error of prediction (RMSEP). If RMSEP is found to be comparable to acceptable root mean square error of calibration, then this indicates that the model is sufficiently accurate when tested with an independent test set. Qualitative applications such as classification, misclassification rate, or positive prediction rate can be used to characterize accuracy.
Precision, with its evaluation of repeatability, intermediate precision, and reproducibility (if greater than one laboratory) are primarily unchanged. Multivariate analysis precision is evaluated with the routine metrics of RMSEP above.
Required for: ID tests, Assay, Purity, and Impurity (quantitative) tests
ProPharma’s team of chemistry, manufacturing, and controls (CMC) experts have over 50 years of cumulative experience and can assist our clients in both reviewing the content of their analytical test method validations per the new expectations as well as authoring the relevant FDA submissions. Our reviews are extremely thorough, going in-depth to make sure that the conclusions listed in the report are evidence-based, data-driven, and will hold up to scientific and regulatory scrutiny.
Interested in learning how we can help with your regulatory and CMC-related regulatory needs? Our team of expert regulatory consultants is here to help.
Contact us today to learn more about our unique approach to working with FDA and our track-record of success with the Agency over the last 40+ years.
TAGS: CMC General Regulatory Regulatory Sciences Analytical Methods
July 16, 2024
The key to successful drug development in the US is directional and focused navigation of FDA’s Investigational New Drug (IND) process. The Chemistry, Manufacturing, and Controls (CMC) section is a...
August 30, 2023
FDA Provides Acceptable Intake Limits for Nitrosamines Drug-Substance Related Impurities (NDSRIs). Can You Comply? The U.S. Food and Drug Administration (FDA) recently issued final guidance entitled;...
November 3, 2015
Issues for Sponsors with INDs and CMC concerns FDA recently released a draft guidance entitled “Recommendations for Microbial Vectors Used for Gene Therapy” as a supplemental guidance to “Guidance...