Calibration and Verification: Are you using these terms correctly?

April 5, 2013

No matter where I go I am always amazed at how many times the words calibration and verification are misused. I often hear statements like, “The instrument doesn’t need to be calibrated, the accuracy just needs to be verified,” or, “The instrument needs to be calibrated to make the measurements more accurate” To many people, the act of calibrating something implies that some adjustment is being made. Actually, the internationally accepted definition of calibration is quite different.

The International Vocabulary of Metrology — Basic and general concepts and associated terms (VIM), JCGM 200, 3rd edition, defines calibration as an operation that:

…under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication

As can be seen by the definition, calibration is the act of comparing values indicated by a standard of known accuracy to corresponding values indicated by the item being calibrated. Nowhere in this definition is adjustment mentioned. In fact, the VIM goes on to state that, “Calibration should not be confused with adjustment of a measuring system, often mistakenly called “self calibration”, nor with verification of calibration.” Verification — a word often misused as a synonym for calibration — is reserved for those activities providing confirmation that an item meets specified requirements. A table listing the calibration status of instrumentation would, for example, be considered a verification of calibration whereas the act of placing a standard mass on an analytical balance to determine the accuracy of the balance would be considered a calibration.

So, if calibration does not involve adjustment, what then is adjustment? Adjustment is just one of many possible corrective actions taken to correct a deficiency in the accuracy of a calibrated instrument. Other corrective actions include application of correction factors, re-assigning a quantity value, or replacement of the inaccurate item. The key term to keep in mind here is corrective action. While some calibrations require a corrective action to restore instrument accuracy, others calibrations do not. Regardless of whether or not a corrective action is taken, if a measured quantity is compared to a quantity of known value, a calibration is being performed.

TAGS:

Using Propagation of Error to Determine Validation Acceptance Criteria for In-Process Measurements

Many times in validation I am confronted with the problem of how to set acceptance criteria for critical process measurements. This is particularly true of in process measurements made during...

How Do You Stay “c”?

Trying to remain compliant with the current Good Manufacturing Practices (cGMP) has its challenges and opportunities. I would like to discuss the opportunity afforded by the ASTM E2500 "Standard...

September 11, 2013

Do Not Neglect GMP in your Serialization Project

In what for some may be a relatively chaotic run-up to meeting California’s 2015 serialization regulation, it is critical for serialization project managers to assure implementation falls within the...